19 - EN
4.3 Stopping the generator
1
Close the ball valve on the N2 Outlet port.
2
Press
to initiate the shutting down sequence.
The generator will complete the current cycle and then exhaust both beds. This may take several minutes
particularly on ppm generators.
3
When the generator is depressurised it will revert to standby mode.
4.4 Menu Interface
All of the operational parameters and data are accessed through the menu driven interface..
The interface will automatically default back to the main operating menu if no key activity has been detected for one minute.
Note:
After an additional two minutes of inactivity the display will darken. To brighten the display press
.
O2 = 5.00%
O2 = 5.00%
Shutting Down
Shutting Down
Shutting Down
Shutting Down
Standby
Standby
O2 = 5.00%
O2 = 5.00%
Shutting Down
Shutting Down
1. Hour meters
1. Hour meters
2. Fault log
2. Fault log
2.1 Fault 10:
2.1 Fault 10:
[---]
[---]
2.1 Fault 1: Active
2.1 Fault 1: Active
C1 at 5234h
C1 at 5234h
3. Customer
3. Customer
settings
settings
3.16 EST enable
3.16 EST enable
= 0
= 0
3.15 Modbus parity
3.15 Modbus parity
= Even
= Even
3.14 Remote control
3.14 Remote control
= 1
= 1
3.13 Machine number
3.13 Machine number
= 1
= 1
3.12 Fault relay
3.12 Fault relay
On Stop =1
On Stop =1
3.11 Economy enable
3.11 Economy enable
= 1
= 1
3.10 Start cycle
3.10 Start cycle
enable = 1
enable = 1
3.9 Oxygen Analogue
3.9 Oxygen Analogue
FSD =6%
FSD =6%
3.8 Oxygen alarm
3.8 Oxygen alarm
delay = 60s
delay = 60s
3.7 Oxygen alarm
3.7 Oxygen alarm
level =15ppm
level =15ppm
3.6 Pressure units
3.6 Pressure units
= Bar
= Bar
3.5 Pressure reading
3.5 Pressure reading
= 9Bar
= 9Bar
3.4 Calibrate EST
3.4 Calibrate EST
sensor
sensor
3.3 Calibrate outlet
3.3 Calibrate outlet
sensor
sensor
3.2 Oxygen Alarm
3.2 Oxygen Alarm
Over-ride =0
Over-ride =0
3.1 Password enable
3.1 Password enable
= 0
= 0
1.4 Hours EST
1.4 Hours EST
4h
4h
1.3 Hours economy
1.3 Hours economy
25h
25h
1.2 Service due at
1.2 Service due at
8000h
8000h
1.1 Hours Run
1.1 Hours Run
5234h
5234h
+
+