4.10. Replacing the control panels/handsets
1. Disconnect the bed from the mains power supply,
2. Remove the middle cover of the base,
3. Unfasten the band clip mounting the control panel/handset cable,
4. Disconnect the control panel/handset from the control box,
5. Connect the new control panel/handset to the electric control box,
6. Fasten the control panel/handset cable by the band clip,
7. Attach the base cover.
4.11. Replacing the scales sensor
Disassembling the scales sensor
1. Disconnect the bed from the mains power supply,
2. Remove the base covers,
3. Unscrew the bolt holding the scales controller cover and remove the cover,
4. Disconnect the cable of the disabled scales sensor,
5. Disconnect the cable from the base frame,
6. Secure the sensor by screwing on the protective element [4],
7. Unscrew the 3 screws fastening the sensor [2],