2.2 Unpacking and first use 

 

Caution!

 

If  the  product  is  installed by  Park  House  Healthcare  it  is  not  the  responsibility  of 

the user to perform the activities described in this item.  

 

 
The bed is shipped by Park House Healthcare assembled and in a wooden, open-work chest. The bed 
should be unpacked indoors to protect from possible damage 
.  
Prior to installation you must: 

-

 

Read the this User Manual carefully 

-

 

Remove all packaging and materials protecting the bed 

 

 

 

Caution! 

Package may be marked with following signs: 

 

 

- multiuse package

 

 

- package for recycling 

 

 

low-density polyethylene packaging

 

    

LDPE 

Package  waste  is  recyclable  and  should  be  segregated  before  removed  to  a  disposal 
site. 
Returnable packaging should be returned to the producer of the medical product

..  

 

-

 

Remove any additional equipment 

-

 

Remove the wheel blocks and release the central locking system  

-

 

With  help of another person carefully remove the bed from the palette and lock the brakes  

 

 

Caution!

 

Do not hold the plastic ABS covers whilst moving the bed  

-

  Ensure the equipment provided fits the documents enclosed 

-

  Ensure all accessories are fitted in accordance with the instructions stated In item

 4 Optional 

elements

 

-

   Install the product in its required destination ensuring the environment meets the 

requirements as described in 

item 1.2 General requirements

 

-

  If  the  place  where  the  bed  has  been  installed  is  equipped  with  earth  grounding  point  you 

should connect the bed with it by means of earth ground wire connected to earth grounding 
socket in the bed connecting panel, 

 

 

Caution!

 

All electric devices connected directly with the patient should be connected to the 
bed earth grounding point 

 

   Connect the bed to the mains ensuring there is suitable electrical supply 
   Check all functions are in correct working prior to patient use 

 
All bed functions should work silently and smoothly, without stopping or creaking etc.  
Any noise produced from the bed and electric servo-motors does not exceed 65dB(A) measured at 1m 
from the bed 
 

Содержание Endeavour

Страница 1: ...Endeavour Bed Service Manual...

Страница 2: ...4 SERVICE 4 1 MAINTENANCE ACTIVITIES 4 2 PRODUCT DISPOSAL 4 3 REPLACING THE BASE COVER 4 4 REPLACING THE SUPPORT COLUMN 4 5 REPLACING THE SERVO MOTOR IN THE BACK THIGH REST SEGMENT 4 6 REPLACING THE C...

Страница 3: ...he bed 2280 20 mm Width with rails 970 20 mm Maximum height 850 20 mm Minimal height 450 20 mm Back rest angle 74 5 Thigh rest angle 45 3 Trendelenburg position 15 3 Anti Trendelenburg position 17 3 C...

Страница 4: ...2 PRODUCT ILLUSTRATION AND FEATURES Fig 1 Pos in fig 1 Description 1 Base 2 Bed frame 3 Back rest section 4 Fixed section 5 Thigh rest section 6 Calf rest section 7 Side rail 8 Foot end Top 9 Castor...

Страница 5: ...edures Back rest and thigh rest sections have double auto regress function The bed is provided with a potential equalising clamp The bed is compatible with all Park House Healthcare pressure relieving...

Страница 6: ...on carefully remove the bed from the palette and lock the brakes Caution Do not hold the plastic ABS covers whilst moving the bed Ensure the equipment provided fits the documents enclosed Ensure all a...

Страница 7: ...rs does not exceed 65dB A measured at 1m from the bed 5 Check all the steering functions of the bed 6 Check all the locking functions on the nurse control panels are operating correctly see descriptio...

Страница 8: ...al inspection and any identified malfunctions should be handled as described in clause 6 6 Location and removing faults 4 2 Product disposal If use of the of the Endeavour bed is no longer required th...

Страница 9: ...ket cover in the electric control box 4 Disconnect the power cord supplying the column from electric control box 5 Cut the insulation grip holding the column cable 6 Support the bed frame from falling...

Страница 10: ...10 Screw in the column to the bed frame with 4 M8 hexagonal bolts 1 11 Screw 3 M6 screws 3 into the column holder 5 12 Fasten the cables connected to the electric control box by the band clip 13 Atta...

Страница 11: ...7 Attach the bed base covers 4 6 Replacing the CPR mechanism cord 1 Disconnect the bed from the mains power supply 2 Disconnect the cord from the servo motor lever Servo motor pull out the plastic sn...

Страница 12: ...from shank section side and hold it in place to prevent it from falling 3 Unscrew M8 screw 3 mounting the gas spring to the bed frame Assembling the gas spring 1 Put on the gas spring 1 fasten it to...

Страница 13: ...the electric control box 1 Screw down the electric control panel to the bed base 2 Connect the servo motor columns and control elements cables to the electric control box according to illustration ab...

Страница 14: ...lectric control box 6 Fasten the control panel handset cable by the band clip 7 Attach the base cover 4 11 Replacing the scales sensor Disassembling the scales sensor 1 Disconnect the bed from the mai...

Страница 15: ...Put the scales controller cover on and screw down the bolts to fasten the cover 8 Attach the base covers 4 13 Replacing the wheel Disassembling the wheel 1 Disconnect the bed from the mains power supp...

Страница 16: ...fers 1 Unscrew the screw 2 fixed to the buffer 1 2 Put a new buffer 1 onto the sleeve 4 3 Put a washer 3 and a sleeve 4 either side the new buffer 1 and screw in place securely 2 4 Screw all elements...

Отзывы: