RINSE
CAUTION
Risk of allergic reactions and irritation of the mucous
membrane by disinfectants
Disinfectants can trigger allergic reactions or irritation of the
mucous membrane on contact with the skin.
• Rinse the product thoroughly to ensure that no residues of
the disinfectant remain on the PARI product.
• Rinse off all parts thoroughly in running drinking water.
Water temperature: approx. 15 °C, duration: 2 minutes
• Dispose of the used solution (the diluted solution can be dis-
posed of down the drain).
DRY:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent surface and al-
low them to dry completely.
Sterilising
CAUTION
Risk of infection by remaining germs
If any germs are left on the parts that are to be sterilised,
they can proliferate despite sterilisation. As a result, there is
a danger of infection.
• Clean, disinfect and dry all components thoroughly before
sterilising.
• Use only validated procedures for cleaning and
disinfection.
EQUIPMENT:
The method has been validated in Europe using:
– Steam steriliser with fractionated pre-vacuum (in accord-
ance with DIN EN 285 or DIN EN 13060 (Type B)
– Sterile barrier system in accordance with DIN EN 11607-1
– 29 –
Содержание O-PEP System
Страница 2: ......
Страница 6: ...Vorreinigen 24 Reinigen und desinfizieren 24 Sterilisieren 29 Visuelle Kontrolle und Aufbewahrung 30 6...
Страница 31: ...31...
Страница 34: ......
Страница 38: ...Precleaning 24 Cleaning and disinfecting 25 Sterilising 29 Visual inspection and storage 30 6...
Страница 63: ...31...
Страница 66: ......
Страница 70: ...Voorbereiden 24 Voorreinigen 24 Reinigen en desinfecteren 25 Steriliseren 30 Visuele controle en opbergen 31 6...