PROCEDURE
• Place all the parts in the cleaning solution.
• Clean the parts thoroughly with a brush.
• Thoroughly rinse the parts under hygienically tested water
and leave them to dry completely on a clean, dry and
absorbent surface.
VALIDATION
The method has been validated using:
– Korsolex
®
Endo Cleaner 0.5% (Bode)
– approx. 50 °C warm water
– Duration of cleaning procedure: approx. 5 minutes
- 37 -
en
Manual disinfection
For manual disinfection, select a suitable instrument disinfectant
that has been proven to be effective and compatible with the
cleaning agent used.
EQUIPMENT
– Disinfectant
– Water
CAUTION
PARI product components are not suitable for treatment with
instrument disinfectants based on
quaternary ammonium
compounds
. Quaternary ammonium compounds tend to
become concentrated in plastic materials, which can cause
incompatibility reactions
in patients.
PARI product components are not ...