- 34 -
5 HYGIENIC RE-USE IN THE PROFESSIONAL
ENVIRONMENT
NOTICE
Ensure that the system is processed for re-use only with the
application of adequately validated device- and product-specific
procedures, and that the validated parameters are maintained
during each cycle. The equipment used must be maintained
regularly.
5.1 General
Ensure that the system is processed ...
Follow the schedule defined below for hygienic re-use:
– Dispose of the used filter pad
after every inhalation session
.
– Clean and disinfect the PARI Filter/Valve Set
after every inha-
lation session
.
– Clean, disinfect and sterilise the PARI Filter/Valve Set
before
every change in patient
.
The PARI Filter/Valve Set must be processed for re-use with the
application of suitable, validated methods in such manner that
the success of these methods can be monitored reliably and the
safety and health of patients is not threatened.
CAUTION
Ensure that the components are allowed to dry sufficiently after
each processing step. Condensation or residual wetness can
present an increased risk through the growth of bacteria.
Ensure that the components are allowed ...
The following text describes the methods validated by PARI, with
which plastic PARI parts can be processed effectively.
Ensure that the desired results are achieved ...
Ensure that the desired results are achieved consistently by the
processing methods employed by your staff with the equipment
and chemicals used. Take particular care to ensure that the pro-
Содержание 041G0500
Страница 1: ...2012 PARI GmbH Spezialisten f r effektive Inhalation 041D0205 D 10 12 Titelseite Filter Ventil Set...
Страница 2: ...Gebrauchsanweisung 3 Instructions for use 23 de en nl...
Страница 4: ...4 SONSTIGES 1 Entsorgung 1 Kontakt 1 ERSATZTEILE UND ZUBEH R 1...
Страница 24: ...24 MISCELLANEOUS 40 Disposal 40 Contact 40 SPARE PARTS AND ACCESSORIES 41...