EQUIPMENT
– Cleaning and disinfection device in conformance with
DIN EN ISO 15883
– Cleaning agent
– Neutralising agent if necessary
PROCEDURE
• Arrange the components so that they can be cleaned most
effectively.
• Select a suitable cleaning programme and agent.
VALIDATION
The method has been validated using:
– G7836 CD cleaner-disinfector produced by Miele,
programme Vario TD.
– Standard baskets manufactured by Miele
– Neodisher Mediclean forte 0.5% and Neodisher Z for neutral-
isation (from Dr. Weigert)
DRYING
Even if the cleaner-disinfector is equipped with a drying func-
tion, make sure that no residual moisture is left in the compo-
nents. If necessary, remove any residual moisture by shaking
the parts and allow them to dry completely.
- 36 -
Mechanical cleaning and disinfection
Manual cleaning
As an alternative to mechanical cleaning and disinfection, the
following manual method can be used:
EQUIPMENT
– Cleaning agent
– Brush
– Warm water