LE 5002 Non Invasive Blood Pressure Meter
USER MANUAL
8
5. EQUIPMENT DESCRIPTION
5.1.
CONTROL UNIT FRONT PANEL
Figure 7. Control Unit Front Panel.
•
CUFF: Pneumatic connection for the tail cuff. It sends air from the internal
pump to the cuff in order to occlude the caudal artery when taking
measurements.
•
TRANSDUCER: DIN 6 connector used to connect pulse transducer.
•
BEATS: Red coloured led that flashes at same frequency that heart beats when
transducer is placed on animal tail.
•
GAIN: Potentiometer used to adjust signal from the transducer.
•
DISPLAY: 4 row, 20 column display. It is used to display the result of
measurements, and several messages depending on the mode the instrument is
in.
•
SEND: Button used to send data to the computer using RS-232 cable and
SEDACOM software. It activates one of 2 options:
o
SEND ACTIVE TRIAL: Sends only the trial shown in the display.
o
SEND ALL MEMORY: Sends all data stored in memory.
•
CLEAR: Button used to erase data stored in the internal memory. It activates
one of 2 options:
o
CLEAR ACTIVE TRIAL: Clears only the trial shown on the display.
o
CLEAR ALL MEMORY: Clears all data stored in memory.
CUFF TRANSDUCER BEATS DISPLAY
GAIN SEND CLEAR ARROWS
START
STOP