LE 5002 Non Invasive Blood Pressure Meter
USER MANUAL
30
12.4.
TRANSDUCER
A transducer in a good state of repair will NOT give pulses (BEAT), when left without
vibrations and at maximum gain (GAIN on the right), although the odd intermittent
impulse may be observed.
12.5.
PROCEDURE TO CLEAN PULSE TRANSDUCERS
To clean the transducers we will use alcohol, talcum powder and a cotton swab. Please
follow these instructions to avoid damaging the transducer during the cleaning
procedure.
Figure 21. Alcohol and talcum powder.
1)
Wet the cotton swab with alcohol.
Figure 22. Wet the cotton with alcohol.
2)
Put the cotton swab in parallel position to the sensor and brush softly by turning
the cotton side. Never brush the sensor by dragging the cotton over it.