LE 5002 Non Invasive Blood Pressure Meter
USER MANUAL
11
5.3.
DISPLAY
Figure 9. Display.
5.3.1. DATA ROW
The first two rows on the display show experiment data. The upper row contains the
labels and the lower one features the numerical values. When an experiment has not
been completed and there are no numerical values the display will show ---.
•
TRI: Is the number of trial. The system automatically increases it after every
experiment. When clearing all the memory it is reset to 1.
•
BPM: Animal’s heart beat frequency in pulses per minute.
•
SYS: Systolic pressure.
•
DIA: Diastolic pressure.
•
MED: Mean pressure calculated with the formula
)
·(
33
.
0
DP
SP
DP
MP
+
+
=
5.3.2. MESSAGE ROW
This line shows several messages depending on the instrument’s current status. They are
listed below:
5.3.2.1. Auto Calibration
The system takes a few seconds to auto calibrate after the instrument is turned on.
5.3.2.2. Pulse level Ready
This message is displayed when the pulse level received by the instrument is suitable
for carrying out a measurement. If this message is shown, press
START to initiate the
measurement process. Otherwise, if you press the aforementioned button the unit will
do nothing and the
GAIN potentiometer should increase.
TRI BPM SYS DIA MED
123 305 167 145 152
<Insufficient level>
□□□□
DATA ROW
MESSAGE ROW
LEVEL ROW