20
This requires the TLD Reader to be calibrated relative to the PSDL or SSDL and the Irradiator exposure adjusted
to this using the same reference TLDs. Please discuss with Panasonic in detail.
For Reference - Automatic Internal Calculation of Source Decay:
Current Dose Rate = ( Initial Dose Rate ) e - ( 0.693 t / T )
t
=
Elapsed time in Days.
T
=
1/2 life of Cs137 in Days.
(30.17 Years x 365.25 Days/Year = 11 019.59 Days)
Set Operator Level Password
- This function changes the operator level password using the mouse and
keyboard.
Set Manager Level Password
- This function changes the manager level password. The manager uses the
mouse and keyboard to change the manager level password.
Irradiator ID Number
- This is the serial number, which is stamped on a label on the back of the unit. This
parameter is normally set by factory personnel. If modification is required, the manager uses the mouse to
highlight "
Irradiator ID Number
", and the new number is then entered.
Parity Checking
- This is a toggle function which alternately enables ("
Enabled
") or disables ("
Disabled
")
automatic parity checking of the badge number code.
Pulse Delay Motor, Hunt Time
- This function allows the pulse delay of the stepper motors to be set. This critical
parameter is normally set by factory personnel to appropriate values. Since proper system operation and error
checking is based on these values, modification by non-factory personnel is not advised.
Number Code Read
– This is a toggle function which alternately enables or disables reading of the badge
number code and it is also available on main screen.
Automatic Batch Create When Program Starts Up
– This option enables automatic batch create feature when
program starts up. The batch file name is in “MMDDYYHH” format where MM is month, DD is date, YY is the last
two digits of year, and HH is military hours.
OK button
- This function saves current settings and returns to the Main screen.
Cancel
- Returns the program to the System Parameter Menu without saving settings.