   ALARMS AND SAFETY FUNCTIONS 

 
The incubator supports the following alarms, safety functions, and self-diagnostic functions. 
 

Alarms and safety functions 

Alarm & 

Safety 

Situation Indication 

Buzzer 

Safety 

operation

Automatic set 
temperature 
alarm 

The internal temperature is over 
+/-2.5 

o

C from the set temp.. 

Alarm lamp lights. 
All digits on the temp.
indicator flash. 

Intermittent tone 

----- 

Key lock   
switch 

When the key lock switch is 
turned ON. 

----- ----- 

Key input is 
disabled. 

Temperature 
sensor 
abnormality 

The temperature sensor is 
disconnected. 

Alarm lamp lights. 
E01 is displayed on 
the temp. indicator. 

Intermittent tone  Heater OFF 

Triac 
abnormality 

The triac is disconnected. 

Alarm lamp lights. 
E02 is displayed on 
the temp. indicator. 

Intermittent tone  Heater OFF 

The triac is short-circuited. 

Alarm lamp lights. 
E03 is displayed on 
the temp. indicator. 

Intermittent tone  Heater OFF 

Relay 
abnormality 

The relay is short-circuited. 

Alarm lamp lights. 
E04 is displayed on 
the temp. indicator. 

Intermittent tone  Heater OFF 

The relay is disconnected or 
the heater is short-circuited. 

Alarm lamp lights. 
E05 is displayed on 
the temp. indicator. 

Intermittent tone  Heater OFF 

Independent 
over-heat 
protection 

(When the unit is not running) 
The safety circuit is activated 
by independent temp. sensor   
if the chamber is being   
overheated abnormally. 

Without display 
change. 

Continuous tone 

Heater OFF 
forcedly by 
external circuit. 

(When the unit is running) 
The safety circuit is activated 
by independent temp. sensor   
if the chamber is being 
overheated abnormally. 

Alarm lamp lights. 
E05 is displayed on 
the temp. indicator. 

Continuous tone 
(When the temp. 
is decreased, 
intermittent tone) 

Heater OFF 
forcedly by 
external circuit. 

* The buzzer tone resulting from the independent over-heat protection cannot be stopped with the alarm 
buzzer stop key (BZ).    Turn off the main switch. 
 
 

    Operation after power failure

 

 
The set value is memorized by nonvolatile memory. Accordingly, the heated incubator resumes the 
operation with setting before power failure. 
 
 
 

Содержание MIR-H163 Series

Страница 1: ...instructions carefully before using this product and save the operating instructions for future use See page 26 for all model numbers MIR H163 MIR H263 Series MIR H263 MIR H163 MIR H263 Heated Incubat...

Страница 2: ...NSTALLATION P 13 PRELIMINARY RUN P 13 KEY OPERATION FLOW P 14 NORMAL MODE AND TIMER MODE C P 15 STABLE RUNNING P 16 STABLE RUNNING WITH DELAY TIME P 17 ALARMS AND SAFETY FUNCTIONS P 18 Operation after...

Страница 3: ...ing instructions are subject to change without notice for improvement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or the pag...

Страница 4: ...in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Symb...

Страница 5: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 6: ...nction Disconnect the power supply plug if there is something wrong with the unit Continued abnormal operation may cause electric shock or fire When removing the plug from the power supply outlet grip...

Страница 7: ...ces such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components or electric parts Check the setting when starting up of operation after power fa...

Страница 8: ...o C Maximum relative humidity 80 for temperature up to 31o C decreasing linearly to 50 relative humidity at 40o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltag...

Страница 9: ...8 INCUBATOR COMPONENTS ON OFF KEYLOCK 8 Outer door 2 1 5 3 4 Inside 6 11 12 7 Shelf 9 10 For model with CE mark...

Страница 10: ...w objects to touch or scratch the sensor 10 Exhaust air vent Do not put anything on this unit as an exhaust air vent may become hot during operation Make sure that the exhaust air vent is not blocked...

Страница 11: ...ered by pressing this key A setting temperature and a setting time of operation timer are displayed Table 1 on next page By re pressing this key the set value lamp goes out and the mode returns to nor...

Страница 12: ...on timer is finished 14 Alarm lamp ALARM This lamp lights when the unit is in warning condition 15 Alarm buzzer stop key BZ Press this key to silence the buzzer in the event that the alarm operates an...

Страница 13: ...cause vibration or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tippin...

Страница 14: ...telephone line or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit PRELIMINARY RUN When using the unit for the first time after purchasing operate the unit wi...

Страница 15: ...re and set operation time is displayed Input the temperature and operation time to set Inputted value is displayed Press the timer mode select key Timer mode C lamp Normal mode turns off Timer mode C...

Страница 16: ...of unit Chamber temperature graph Image When a delay timer being set When an operation timer being set During continuous operation During stop operation Start lamp Delay timer lamp Digital temperature...

Страница 17: ...ation sequence Example Chamber temperature 37 o C Continuous running Description of operation Key operated Indication after operation 1 Press the power key The current chamber temperature is displayed...

Страница 18: ...The left digit of the temperature indicator flashes 7 Set the temp to 50 0 and the running time to 01 00 with the digit shift key and numeric value shift key 8 Press the timer mode key to select the t...

Страница 19: ...circuited Alarm lamp lights E04 is displayed on the temp indicator Intermittent tone Heater OFF The relay is disconnected or the heater is short circuited Alarm lamp lights E05 is displayed on the tem...

Страница 20: ...cloth damped with neutral detergent Afterwards wipe off with a cloth washed in clean water Remove the separation plate at the bottom of the oven and wipe off any particles in the heater box using a so...

Страница 21: ...ration Has the set temperature value been changed Has the door been left open for a long time Has a low temperature object been placed in the unit In this case the alarm will be automatically cleared...

Страница 22: ...ed in an unsupervised area for an extended period ensure that children do not have access and doors cannot be closed completely The disposal of the unit should be undertaken by appropriate personnel A...

Страница 23: ...ase help us to conserve the environment we live in German Ihr Panasonic Produkt wurde entworfen und hergestellt mit qualitativ hochwertigen Materialien und Komponenten die recycelt und wiederverwendet...

Страница 24: ...e mani re inappropri e peuvent s av rer potentiellement dangereuses pour la sant humaine et pour l environnement Cependant ces mati res sont n cessaires au bon fonctionnement de votre appareil ou de v...

Страница 25: ...istemas de recolha separados para produtos el ctricos e electr nicos usados Por favor ajude nos a conservar o ambiente em que vivemos Italian Il vostro prodotto Panasonic stato costruito da materiali...

Страница 26: ...he en elektronische apparatuur bij de daarvoor wettelijke instanties aangeboden Alstublieft help allen mee om het milieu te beschermen Swedish Din Panasonic produkt r designad och tillverkad av materi...

Страница 27: ...kg 61 kg Note Design or specifications will be subject to change without notice PERFORMANCE Product name Heated Incubator MIR H163 Model No MIR H163 PA MIR H163 PK MIR H163 PE MIR H163 PU Temperature...

Страница 28: ...on of the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on b There is some danger see below P...

Страница 29: ...Printed in Indonesia 7FB6P15175600 S1014 0 1 1 1 Sakata Oizumi Machi Ora Gun Gunma 370 0596 Japan Panasonic Healthcare Co Ltd 2014...

Отзывы: