16
STABLE RUNNING
Table 4 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
temperature is 37
o
C and the operation status is a continuous running.
Note:
The initial setting, which are set at factory,
are that the delay timer is OFF, the chamber
temperature is 0
o
C, the running time is ever ON,
the timer mode is the normal mode.
Table 4: Basic operation sequence (Example: Chamber temperature; 37
o
C, Continuous running)
Description of operation
Key operated
Indication after operation
1 Press the power key.
The current chamber temperature is displayed.
2 Press the delay timer key.
The current delay time is displayed.
3
Press the delay timer key
to reset the delay time.
Reset the delay time.
4 Press the delay timer key
The current chamber temperature is displayed.
5 Press the call key.
The left digit of the temperature indicator flashes.
6
Set the temp. to 37.0
with the digit shift key
and the numeric value shift key.
9 Press the run/stop key.
The setting mode is completed and the unit runs.
POWER
DELAY
DELAY
DELAY
CALL
RUN
STOP
Time
37
o
C