![Panasonic MDF-U33V Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/panasonic/mdf-u33v/mdf-u33v_specifications_267964012.webp)
-86°C Ultra-Low Freezers
Description
Model Number
MDF-U33V-PA
MDF-U53VA-PA
MDF-U56VC-PA
MDF-U76VC-PA
MDF-U76VA-PA
Interior Volume
11.8 cu.ft.
334 L
18.3 cu.ft.
519 L
18.6 cu.ft.
526 L
25.7 cu.ft.
728 L
25.7 cu.ft.
728 L
Area Footprint, Nominal
6.25 sq.ft.
0.58 m
2
7.24 sq.ft.
0.68 m
2
7.24 sq.ft.
0.68 m
2
9.51 sq.ft.
0.88 m
2
9.51 sq.ft.
0.88 m
2
Electrical, 60 Hz
115V, AC, 15 amp
115V, AC, 20 amp
208/230V, AC, 15 amp
208/230V, AC, 15 amp
115V, AC, 20 amp
Patented VIP
®
Construction
Insulated Inner Door
Steel-framed, high impact plastic with foam-in place insulation
Inner Door Configuration
2
2 (4 optional)
2 (4 optional)
2 (4 optional)
2 (4 optional)
Door Gaskets
Multi-point compression gaskets
Condenser Filter, Easy Cleaning
Electrostatic filter standard, front accessible for easy access, no tools required
Rounded Interior Corners
Simplify Cleaning
Shelf Brackets
Incrementally adjustable
Casters
High-impact 4 with levelling feet
Seismic Restraints
Standard; hard connection to facility wall
Access Ports
Multiple ports allow use of injection tubes, probes, leads, instrumentation
Dimensions & Weight
Interior (w x f-b x h)
19.3” x 23.6” x 44.9”
490 x 600 x 1140 mm
24.8” x 23.6” x 54.3”
630 x 600 x 1380 mm
24.8” x 23.6” x 55.1”
630 x 600 x 1400 mm
34.2” x 23.6” x 55.1”
870 x 600 x 1400 mm
34.2” x 23.6” x 55.1”
870 x 600 x 1400 mm
Exterior (w x f-b x h)
26.4” x 34.1” x 73.2”
670 x 867 x 1860 mm
30.3” x 34.4” x 78.3”
770 x 870 x 1990 mm
30.3” x 34.4” x 78.3”
770 x 870 x 1990 mm
39.8” x 34.4” x 78.3”
1010 x 870 x 1990 mm
39.8” x 34.4” x 78.3”
1010 x 870 x 1990 mm
Net Weight, Empty
562 lbs
255 kg
660 lbs
299 kg
672 lbs
305 kg
805 lbs
365 kg
820 lbs
373 kg
Refrigeration System, HFC Refrigerants, Panasonic Cool Safe High Performance Compressors
High Stage
1 HP
½ HP
1 HP
1 HP
1 HP
Low Stage
¾ HP
1 HP
1½ HP
1½ HP
1½ HP
Voltage Booster, Built-In
Not Available
Not Available
Standard
Standard
Not Applicable
Sound Abatement
Standard
Standard
Standard
Standard
Standard
Refrigeration System, HFC Refrigerants, Panasonic Cool Safe High Performance Compressors
Power, AC, 1 Phase
115V
115V
208-230V
208-230V
115V
Recommended Breaker, Dedicated
15 amp
20 amp
15 amp
15 amp
20 amp
NEMA Plug / Receptacle
5-15P/5-15R
5-20P/5-20R
6-15P/6-15R
6-15P/6-15R
5-20P/5-20R
VIP
®
Series Specifications
Product conforms to RoHS
(European Restriction of Hazardous
Substance directives)
Panasonic Healthcare Company of North America
1300 Michael Drive, Suite A, Wood Dale, IL 60191
Toll Free USA (800) 858-8442, Fax (630) 238-0074
www.panasonic.com/biomedical
© Panasonic Printed in USA 2012.02 JB