Chapter 2 I/O terminal block
2-12
2
2-3-5
HL-D3SMI Operation During Measurement
When send button "SEND" for setting is pressed
Immediately after
"SEND"
is performed, the sensor stops measurement temporarily.
After restart of continuous measurement, the data is unfixed (all output OFF) for
the period of sampling cycle x number of average movements.
When image load button "Get image" is pressed
When image is loaded, measurement of the sensor stops temporarily and the
measurement is then restarted.
Stop of measurement
70ms
~
200ms
ON
OFF
Measurement load
command
"Get Image"
Internal measurement
Continuous measurement
Start of continuous measurement
Get image
Image loading
Calculation
measurement result
Retention of measurement value
Measurement value output after change of setting
*6
Retention of measurement value
Data unfixed
Stop of measurement
Output of measurement
70ms
~
200ms
ON
OFF
Send command
"SEND" for setting
Internal measurement
*5
Continuous measurement
Start of continuous measurement
Retention of
measurement value
Data unfixed
Calculation
measurement result
*5 Sampling cycle x Number of average movements
*6 Sampling cycle x Number of average movements
Continuous measurement
Output of measurement
Measurement value output after change of setting
Continuous measurement
Measurement value output after change of setting
Retention of measurement value
Output of measurement value