Sentinel LNG User’s Manual
57
Chapter 4. Advanced Programming
Chapter 4. Advanced Programming
Note:
The Advanced Program is accessible only to persons with Field Service, Factory or Engineering status.
The Advanced Program enables the entry of parameters in the following categories:
•
Transducer - enter data for preprogrammed or special clamp-on transducers
•
K Factor - enter the Meter Correction (K) Factor as a single value or as a table of values
•
Pipe - enter pipe parameters
•
Weight Factors - enter the appropriate weight parameter.
To measure flow rate with the
Sentinel LNG
, activate the channel/path(s) and enter the transducer, pipe and fluid
parameters. Appendix A includes a complete set of menu maps for the
User Program
, including Figure 231 on page 212
for the
PROG
menu.
4.1 Entering Transducer Parameters
To enter parameters for preprogrammed or special clamp-on transducers, use the
Transducer
menu. Remember to
record all programmed data in Appendix C,
Data Records.
To access the
Transducer
submenu, from the main screen:
1.
Press
[ESC]
. The
Sentinel LNG
enters the
User Program
.
2.
Press the [
] key until
PROG
is bracketed in the top left corner and press
[ENT]
.
3.
Use the [
] and [
] keys to scroll to the desired Channel and press
[ENT]
.
4.
Scroll to the
Transducer
submenu and press
[ENT]. Clamp-on and Wetted options appear.
4.1.1
Clamp-On Transducers
1.
To set up
Clamp-on
transducers, scroll to
Clamp-on
and press
[ENT].
2.
Scroll to either
Preprogrammed
(for the standard transducers) or
Other
(for special transducers), and press
[ENT].
3.
The program also calls for the Wedge Temperature
.
Scroll to the
Wedge TMP
option and press
[ENT].
Then use the
[
] and [
] keys to enter the temperature, and press
[ENT].
Note:
The wedge temperature of the transducer can be approximated by inputting an average value for the
surface temperature of the outside pipe wall.
4.
Do one of the following:
• For preprogrammed transducers, scroll to the desired Transducer Number (407 [2 MHz], 408 [4 MHz], 409
[500 kHz] or 410 [1 MHz]) and press
[ENT].
Then press
[ESC]
three times to return to the
PROG
menu.
• For other
Clamp-on
transducers, proceed to
Other Transducers
on the next page.
IMPORTANT:
Other (special) transducers have no engraved number on the housing and are rarely used. Examine the
transducer housing carefully for a number.
4.1.2
Other Transducers
1.
The first required parameter is the Frequency. Press
[ENT]
to open the
Frequency
window. Then scroll to the
frequency of your transducer (from 0.25 to 4.00 MHz) and press
[ENT]
.
2.
The meter next asks for the Time Delay
(Tw)
.
Scroll to the
Tw
option and press
[ENT].
Then use the arrow keys to
enter the time provided by Panametrics Sensing (in microseconds), and press
[ENT].
3.
The next parameter is the Wedge Angle, the angle of the transducer’s ultrasonic transmission in the transducer
wedge. Scroll to the
Wedge Ang
option and press
[ENT].
Use the arrow keys to enter the provided angle (in
degrees), and press
[ENT].
4.
To enter the Wedge Soundspeed,
scroll to the
Wedge SS
option and press
[ENT]
. Use the arrow keys to enter the
provided soundspeed (in m/s or ft/s), and press
[ENT].
5.
The final parameter is the Temperature Coefficient. Scroll to the
TempCo
option and press
[ENT].
Use the arrow
keys to enter the provided coefficient, and press
[ENT].
Set the coefficient to 0 if you are unsure of the value.
Содержание Sentinel LCT
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Preface iv Sentinel LNG User s Manual...
Страница 12: ...Contents x Sentinel LNG User s Manual...
Страница 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Страница 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Страница 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Страница 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Страница 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Страница 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Страница 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Страница 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Страница 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Страница 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Страница 244: ...Index 232 Sentinel LNG User s Manual...
Страница 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Страница 247: ......