Chapter 5. Using PanaView
152
Sentinel LNG User’s Manual
5.9.4.2
RTD Input Card (cont.)
7.
To calibrate the RTD Input, click on
CAL
, and the following screen appears.
Figure 196: RTD Input Calibration Options
8.
Enter the
Set Pt
or
Offset
temperature into each text box, click on
BACK
, and confirm the value by clicking
Yes
on the
Verification
screen. Click on
BACK
to return to the
RTD Input Menu
.
9.
Click on
Reset Table
and the following screen appears.
Figure 197: Reset Table Confirmation
10.
Click
Yes
twice to confirm the resets, or
No
to reject them. The screen returns to the
RTD Input Menu
.
Содержание Sentinel LCT
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Preface iv Sentinel LNG User s Manual...
Страница 12: ...Contents x Sentinel LNG User s Manual...
Страница 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Страница 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Страница 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Страница 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Страница 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Страница 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Страница 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Страница 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Страница 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Страница 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Страница 244: ...Index 232 Sentinel LNG User s Manual...
Страница 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Страница 247: ......