Sentinel LNG User’s Manual
49
Chapter 3. General Programming
3.7.4 Setting Up API
In the API Setup submenu, set the table types and conditions for several API options:
1.
If the API option is included, from the COMPOSITE menu, select API Setup, and press [ENT]. The next screen displays
API Table Type and API Conditions.
3.7.4.1
API Table Type
1.
Select API Table Type and press [ENT]. The next screen displays four table types: Table A-Crude or JP, Table B-Gen
Product, Table C-User Def TE, and Table D-Lubricants.
2.
For Tables A, B and D, select a specific table, press [ENT], and the screen returns to the API Setup menu.
3.
Select Table C, press [ENT], and the Thermal Exp Coeff screen appears. To change the coefficient, press [ENT]. On
the next screen, use the arrow keys to change the digits, and press [ENT] to save the change, or [ESC] to reject
the change. The screen returns to the previous display. Press [ESC] twice to return to the API Setup display.
3.7.4.2
API Conditions
1.
Select API Conditions. Press [ENT]. The next screen displays two options: Base Condition and Flow Measure At...
2.
Select Base Condition and press [ENT]. The next screen displays four options: 60°, 20°, 15° and User Defined...
Select a given temperature or define a new one.
3.
To establish a new base temperature, select User Defined, press [ENT], and the User Def Base Temp screen
appears. Press [ENT], and the User Defined temperature is able to be changed.
4.
Using the arrow keys, change the appropriate digits, and press [ENT]. Press [ESC] twice and return to the APIcond
screen.
5.
Select Flow Measure At... and press [ENT]. A screen appears listing two options: Observed and Alternate...
6.
To measure at Observed, select Observed and press [ENT]. Two settings appear: Temp(alt) and Pres(alt).
7.
To change either alternate condition, select the option, and press [ENT].
8.
Using the arrow keys, change the appropriate digits, and press [ENT].
9.
Change the other alternate condition if needed, and press [ESC] to return to the Flow Measurement At... screen.
10.
To measure at Alternate, select Alternate and press [ENT]. Three settings appear: Temp(obs), Pres(obs), and
Dens(obs).
11.
To change an observed condition, select the option, and press [ENT].
12.
Using the arrow keys, change the appropriate digits, and press [ENT].
13.
Change the other observed conditions if needed, and press [ESC] four times to return to the PROG menu.
Содержание Sentinel LCT
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Preface iv Sentinel LNG User s Manual...
Страница 12: ...Contents x Sentinel LNG User s Manual...
Страница 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Страница 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Страница 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Страница 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Страница 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Страница 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Страница 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Страница 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Страница 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Страница 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Страница 244: ...Index 232 Sentinel LNG User s Manual...
Страница 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Страница 247: ......