DigitalFlow™ GM868 Service Manual
21
Chapter 2. Error Codes and Screen Messages
2.18 Screen Messages
A variety of messages may appear on the display screen during the performance of a task. Since the error codes
have already been discussed in this chapter and the locator bar messages are discussed in detail in Chapter 3,
Operation, of the Startup Guide, they will not be repeated here. All other messages are listed in Table 2 below.
Table 2: Screen Messages
Message
Meaning
Power Up Messages
Backup Battery FAIL
The backup battery that powers the non-volatile RAM has failed.
Contact the factory.
Backup Battery OK
The backup battery that powers the non-volatile RAM has passed.
DSP Processor OK
The DSP (digital signal processor) has passed.
DSP RAM Failed
The DSP (digital signal processor) RAM has failed. Contact
the factory.
DSP RAM OK
The DSP (digital signal processor) RAM has passed.
EPROM sum = XXXX
Record the EPROM sum at the initial power up and periodically
thereafter.
FRIGID_INIT Executed
The NVR (non-volatile RAM) was automatically initialized due to a
memory fault. Contact the factory.
NVR FAIL
The non-volatile RAM has failed. Contact the factory.
NVR OK
The non-volatile RAM has passed.
Measurement Mode Messages
All Logs Cleared! hit key
This message may appear during one of the following tasks:
1. clearing a log - there are no more logs to clear
2. logging data - user has hit
[ENT]
instead of selecting an old log
or entering a new name
3. printing a log - there are no logs to print
4. displaying a log - there are no logs to display
Hit any key to resume taking measurements.
All Sites Cleared! hit key
This message may appear during one of the following tasks:
1. saving a new site - a new site name was not entered or
overwrite existing site was not selected
2. recalling a site - there are no site files to recall
3. clearing a site - there are no sites to delete
4. printing a site - there are no sites to print
5. displaying a site - there are no sites to display
Hit any key to resume taking measurements.
Do you want to SAVE?
This message appears upon exiting the
User Program, if the new
site data has not been saved. Failure to save will result in loss of
the new data next time the site data is changed or recalled.
Duplicate name,
Enter another.
The site file or log name is already in use. Enter a different name.
End Time must exceed
Start Time by 5 min.
This message appears when in the
LOG
menu. Enter an end time
that is at least five minutes later than the start time.
Header invalid
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Log Active, END only
hit any key
When in the
LOG
menu, this message indicates that the log is still
compiling data. Only the End Time may be edited.
Содержание DigitalFlow GM868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 9: ...DigitalFlow GM868 Service Manual vii Contents...
Страница 10: ...Contents viii DigitalFlow GM868 Service Manual...
Страница 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Страница 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 48: ...Chapter 4 Parts Replacement 38 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 60: ...Appendix B Optional Enclosures 50 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Страница 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 71: ......