DigitalFlow™ GM868 Service Manual
11
Chapter 1. Calibration
1.7
Testing the Alarm Relays
Alarm relays may be added to the Model GM868 by installing an
Alarms Option Card in one (or more) of the six
expansion slots. Each option card includes three alarm relays, which are designated as A, B, and C. To test the alarm
relays, connect an ohmmeter to the option card terminal block as shown in Figure 4 below. Refer to the menu map in
Figure 6 on page 15 as a guide.
Figure 4: Typical Ohmmeter Connections
For this discussion, assume that the option card has been installed in
Slot x
.
1.7.1
Accessing the Calibration Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate
the desired slot
. (The option bar will include a slot listing for each installed option card.)
3.
Press
[F1]-[F3]
to select alarm relay A, B or C, respectively.
Terminal Block (Option Card)
A
COM
A
B
A
NO
C
C
B
B
-
+
Ohmmeter
1
9
C
NC
COM
COM
NO
NC
NO
NC
Содержание DigitalFlow GM868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 9: ...DigitalFlow GM868 Service Manual vii Contents...
Страница 10: ...Contents viii DigitalFlow GM868 Service Manual...
Страница 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Страница 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 48: ...Chapter 4 Parts Replacement 38 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 60: ...Appendix B Optional Enclosures 50 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Страница 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 71: ......