DigitalFlow™ GM868 Service Manual
19
Chapter 2. Error Codes and Screen Messages
2.6 E4: Signal Quality
Problem:
The signal quality is outside the limits programmed in the
SETUP
submenu of the
User Program.
Cause:
Too high a signal strength may be caused by the failure of an electronic component. Too low a signal
strength may be caused by a flowcell or electrical problem.
Action:
Check for sources of electrical interference and verify the integrity of the electronics console by
temporarily substituting a test flowcell that is known to be good. Check the transducers and relocate
them, if necessary. See Chapter 3,
Diagnostics, for instructions.
2.7 E5: Amplitude Error
Problem:
The signal amplitude exceeds the limits programmed in the
SETUP
submenu of the
User Program.
Cause:
Excessive levels of an attenuating gas, such as CO
2
, are present in the flowcell. Solid or liquid particulates
may be present in the flowcell.
Action:
Refer to Chapter 3,
Diagnostics, to correct any flowcell problems.
2.8 E6: Cycle Skip, Accel.
Problem:
The acceleration exceeds the limits programmed in the
SETUP
submenu of the
User Program.
Cause:
This condition is usually caused by poor flow conditions or improper transducer spacing.
Action:
Refer to Chapter 3,
Diagnostics, to correct any flowcell and/or transducer problems.
2.9 E7: Analog Out Error
Problem:
The current in the analog output circuit exceeds the limits for the analog output port.
Cause:
The output load exceeds the specified limits for the analog output port.
Action:
Make sure the output load is <550 ohms for the
Slot 0
analog outputs or is <1000 ohms for an analog
outputs option card (slots 1-6).
2.10 E8: Temp In
Problem:
This message indicates a temperature input error.
Cause:
The temperature exceeds the specified limits for the analog inputs option card.
Action:
Check the temperature transmitter and the connecting cable. Refer to Chapter 1,
Calibration, and
recalibrate the analog inputs on the option card.
2.11 E9: Press In
Problem:
This message indicates a pressure input error.
Cause:
The pressure exceeds the specified limits for the analog inputs option card.
Action:
Check the pressure transmitter and the connecting cable. Refer to Chapter 1,
Calibration, and recalibrate
the analog inputs on the option card.
2.12 E10: Special Input
Problem:
This message indicates a special input error.
Cause:
The special input exceeds the specified limits for the analog inputs option card.
Action:
Check the special transmitter and the connecting cable. Refer to Chapter 1,
Calibration, and recalibrate
the analog inputs on the option card.
2.13 E11: Not Used
This error code message is not currently in use.
Содержание DigitalFlow GM868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 9: ...DigitalFlow GM868 Service Manual vii Contents...
Страница 10: ...Contents viii DigitalFlow GM868 Service Manual...
Страница 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Страница 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 48: ...Chapter 4 Parts Replacement 38 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 60: ...Appendix B Optional Enclosures 50 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Страница 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 71: ......