Chapter 1. Calibration
4
DigitalFlow™ GM868 Service Manual
1.4 Accessing the Calibration Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate
the desired slot
. (The option bar will include a slot listing for each installed option card.)
3.
Press
[Fx]
to select the desired output.
IMPORTANT:
The procedure for calibrating
all the outputs is the same
. However, when calibrating
a different output
,
be sure to reconnect the ammeter to the appropriate pins on terminal block
I/O
. See Figure 1 on page 2
or Figure 2 above for the correct pin numbers.
1.4.1
Calibrating the Low End of the Output Range
1.
Press
[F1]
to calibrate the low end of the output range.
2.
Press
[F1]
or
[F2]
to adjust the ammeter reading
UP
or
DOWN
, until a 4 mA reading is achieved. If you press
[F3],
Numer,
enter a numeric value for the calibration point and press
[ENT].
3.
Press
[F4]
to
STORE
the setting, or press
and
[F1]
to
ABORT
the calibration.
Note:
If the ammeter reading cannot be adjusted within 5.0
A of the 4 mA setting, contact the factory for
assistance.
1.4.2
Calibrating the High End of the Output Range
1.
Press
[F2]
to calibrate the high end of the output range.
2.
Press
[F1]
or
[F2]
to adjust the ammeter reading
UP
or
DOWN
, until a 20 mA reading is achieved. If you press
[F3],
Numer,
enter a numeric value for the calibration point and press
[ENT].
Note:
If the ammeter reading cannot be adjusted within 5.0
A of the 20 mA setting, contact the factory for
assistance.
3.
Press
[F4]
to
STORE
the setting, or press
and
[F1]
to
ABORT
the calibration.
Note:
Skip the next section if the linearity test will not be performed at this time. Proceed to Procedure Options on
the next page.
1.4.3
Testing the Analog Output Linearity
Note:
If you exited this menu and are returning, follow steps 1 through 3 in Accessing the Calibration Menu on
1.
Press
[F3]
to
TEST
the linearity of the currently selected analog output.
2.
Check the ammeter reading at the 50% output level.
3.
Then, enter a different output level (0-100%) and press
[ENT]
.
4.
Check the ammeter reading at this setting and press
[ENT]
when done.
Table 1 on the next page lists the expected ammeter readings at various
% Full Scale settings, for both 4-20 mA and
0-20 mA scales. Refer to this table to verify the accuracy of the ammeter readings taken above.
Содержание DigitalFlow GM868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 9: ...DigitalFlow GM868 Service Manual vii Contents...
Страница 10: ...Contents viii DigitalFlow GM868 Service Manual...
Страница 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Страница 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 48: ...Chapter 4 Parts Replacement 38 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 60: ...Appendix B Optional Enclosures 50 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Страница 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Страница 71: ......