Chapter 1. Calibration
14
DigitalFlow™ GM868 Service Manual
1.8.3
Entering Pulse Number
5.
Enter a new frequency in the range of 1-10,000 Hz and press the
[ENT]
key. Verify that the frequency counter reads
the correct value.
6.
Repeat Steps 3, 4 and 5 to test all four of the frequency/totalizer outputs. If any of the outputs fails to pass the
test, contact the factory for assistance.
Procedure Options
You have completed testing the totalizer/frequency outputs. Do one of the following:
• To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
• To leave the Calibration Menu, press
[F4] (EXIT)
twice.