![Panametrics C-ET Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/panametrics/c-et/c-et_installation-manual_3403828034.webp)
26
Model C-ET Installation Guide
1.3.4.2
Preparing the Pipe
1.
Remove any pipe insulation from the installation site.
Note:
Make sure you clear an area that is large enough for the transducers’ installation.
2.
Remove any loose paint or rust with sandpaper. Make sure to preserve the original curvature of the pipe. If the
finish is mirror-smooth, use the file to roughen the surface.
1.3.4.3
Surveying the Pipe
1.
Measure the pipe circumference.
IMPORTANT:
Do not use a calculated value or a nominal value for the circumference.
2.
Mark a line around the entire circumference of the pipe.
Side View
Circumferential Line