16
Model C-ET Installation Guide
1.3.3.2
Setting the Transducer Spacing (cont.)
3.
Slide the other mounting block to the calculated spacing
plus
2 in. (5 cm). For example:
a.
Spacing for the first mounting block = 2 in. (5 cm)
b.
Spacing as calculated by the flowmeter = 0.5 in. (12.5 mm)
c.
Second mounting block final location = 2 in. + 0.5 in. = 2.5 in. (5 cm + 1.25 cm = 6.25 cm)
The overall spacing between blocks should be left edge to left edge, or right edge to right edge.
Note:
If possible, the second block should also be set at least 2 in. (5 cm) from the edge of the fixture.