Chapter 5. Specifications
42
BWT™ System Installation Guide
5.3 Threaded Buffer Assemblies
Service
Liquids
Mounting
1 in. NPT
Materials
•
Standard: 316L stainless steel
•
Optional: Titanium
Fluid Temperature
•
FSPA Short Buffers: -40°F to 212°F (-40°C to 100°C)
•
FSPA Extended Buffers: -40°F to 600°F (-40°C to 315°C)
5.4 Socket-Weld Buffer Assemblies
Service
Liquids
Mounting
1 in. socket weld
Material
Standard: 316L stainless steel
Fluid Temperature
•
FSPA Short Buffers: -40°F to 212°F (-40°C to 100°C)
5.5 System
Area Classifications
•
Explosion-proof: Class I, Division 1, Groups C&D
•
Flameproof:
BWT:
II 2 G Ex d IIC T6 Gb KEMA 01ATEX2051
Ex d IIC T6 Gb IECEx KEM 09.0010
JB:
II 2 G Ex d IIC T6 Gb DEKRA 13ATEX0120X
Ex d IIC T6 Gb IECEx DEK 13.0034X
European Compliance
Complies with directives 2004/108/EC EMC, 2006/95/EC LVD (Installation Category II, Pollution Degree 2),
94/9/EC ATEX and 97/23/EC PED for DN<25.
Pipe Sizes
2 in. to 30 in. (50 mm to 750 mm), larger sizes upon request
Velocity Ranges
•
Gas service: 0.1 to 150 ft/s (0.03 to 46 m/s)
•
Liquid service: 0.1 to 40 ft/s (0.03 to 12 m/s)
Note:
The maximum flow velocity specification for gases is variable, depending on gas sound speed, ultrasonic
path length and gas density (pressure and molecular weight).
Содержание BWT System
Страница 1: ...panametrics com Flow 916 058 F July 2021 BWT System Installation Guide ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 8: ...Contents BWT System Installation Guide ...
Страница 16: ...Chapter 1 General Information 8 BWT System Installation Guide no content intended for this page ...
Страница 52: ...Chapter 5 Specifications 44 BWT System Installation Guide no content intended for this page ...
Страница 54: ...46 BWT System Installation Guide no content intended for this page ...
Страница 57: ......