Chapter 3. Acoustic Isolation Installation (FIPA Buffers)
32
BWT™ System Installation Guide
IMPORTANT:
In the next step,
DO NOT
overtighten the transducer.
3.
Torque the transducer to a maximum of 15-18 ft-lb (20- 25 N-m).
4.
Verify that the junction box is properly oriented as discussed in the previous section (see
5.
To connect the transducer electrical cables, see the
Startup or User’s Manual that was supplied with your
flowmeter.
6.
After you install the FIPA Buffers and transducers, you must allow the epoxy to cure for 8 hours. However, while the
epoxy cures, you should verify that the transducer and buffer are working properly (see the next section) below.
7.
Repeat steps 1-6 to install the other transducer.
CAUTION!
Do not place insulation on or around the transducer or junction box. The transducer and
junction box act as a heat sink that protects the transducer from high and low
temperatures.
CAUTION!
While the epoxy cures, do not remove, re-torque or adjust the transducer or you will
crack the epoxy.
Top View
BWT Transducer/Junction Box