33
4. Adjust auto logout time
•
Login to the system as supervisor
•
Press the setup button on the main menu screen
•
Select the ‘auto logout’ function using the up and down arrows
•
Press the curved arrow to go to the entry screen
•
Enter the time for auto logout then press enter
5. View runtime counter indicators for system runtime and motor runtime
•
Login to the system as supervisor
•
Press the setup button on the main menu screen
•
Select the ‘runtime hrs’ function using the up and down arrows
•
Press the curved arrow to go to the view screen
•
The screen will indicate accumulated runtime separately for both system and motor
rotation
•
Press
to return to the settings list magnetic mixer
6. Export of Recipe library to memory card
•
Insert a micro SD card into the memory slot (See section 0)
•
Login to the system as supervisor
•
Press the setup button on the main menu screen
•
Select the ‘recipe imp/exp’ function using the up and down arrows
•
Press the curved arrow to open the next screen (Figure 36)
•
Press the file number button to go to the entry screen
•
On the entry screen input the file number (0-9) you wish to export and press enter
•
Press the export recipes button and confirm when prompted (Figure 37).
•
The entire library will be exported to the micro SD card under filename
LIBRARYX.BAK, where X is the file number entered on the previous step.
•
A confirmation notice will appear after the file writing is finished (this may take several
minutes)
•
Acknowledge the notice and press
to return to the settings list
Figure 36: Import/Export screen (left). Confirmation screen (right)
Содержание Allegro MMG403
Страница 39: ...39 Figure 38 Left Standard configuration Right Extended configuration ...
Страница 47: ...47 15 Electrical Diagram Figure 48 Electrical diagram control box ...
Страница 48: ...48 Figure 49 Electrical diagram drive box ...
Страница 50: ...50 16 CE Certificate ...
Страница 53: ...53 ...