15
2. Verify the correct article code and expiration date. Remove the outer and inner packaging
bag by tearing open at the easy to tear notch (Figure 14) - DO NOT USE SCISSORS OR
OTHER CUTTING IMPLEMENTS. Retain the label from the inner packaging according to
the quality policies of your organization.
3. Visually inspect the Allegro mixing biocontainer for shipping damage or imperfections;
cuts, tears or punctures; film cracks that are externally rough/sharp to the touch (note:
white creases that are not sharp to the touch are typically not a cause for concern);
impeller detached from impeller seat/post; kinked tubing that cannot be unkinked;
damaged, missing or detached subcomponents. If any such imperfections are discovered,
contact Pall or the user organization’s quality group for advice on whether the Allegro
mixing biocontainer should be used.
4. A blue shipping disc is magnetically attached to the outside of the Allegro mixing
biocontainer, over the impeller seat. Reverse the Allegro mixing biocontainer to have
access to the magnetic impeller with the blue disc. Next, remove the blue shipping disc
(Figure 15) (retain this item; this will be used to secure the impeller during subsequent
Allegro mixing biocontainer disposal).
Figure 14: Removing blue shipping disc from Allegro mixing biocontainer
Figure 13: Opening packaging bags via the easy tear notch