8
Specification/Features
25
3
0 to 20.00 (0.01) µS/cm
20.0 to 200.0 (0.1) µS/cm
200 to 2000 (1) µS/cm
2.01 to 20.00 (0.01) mS/cm
20.1 to 200.0 (0.1) mS/cm
Conductivity Ranges
(Resolution)
3
0 to 10.00 (0.01) ppm
10.0 to 100.0 (0.1) ppm
100 to 1000 (1) ppm
1.01 to 20.00 (0.01) ppt
20.1 up to 200.0* (0.1) ppt
*depending on TDS factor used
TDS Ranges (Resolution)
0.10 to 5.00%
0.01%
Salinity % Resolution
1.0 to 50.0 ppt
0.1 ppt
Salinity ppt Resolution
3
3
±1% full scale
Accuracy
3
3
-10.0 to 110.0ºC
0.1ºC
±0.5ºC
Temperature
Resolution
Accuracy
0.1, 1, 10 0.1, 1, 10
Selectable
Cell Constant
3
3
Automatic / Manual (from 0 to 50 °C)
Temperature Compensation
3
3
3
3
0.0 to 3.0% per °C
25.0 , 20.0 °C (selectable)
Temperature Coefficient
Normalisation Temperature
3
3
0.4 to 1.0
TDS Factor
5
5
Maximum 1 per range
Calibration points
3
3
Selectable
Auto - or Manual-ranging
3
Selectable
Auto standard recognition
TDS
COND
Features
Auto-Buffer Recognition
Yes
Hold Function
Yes “HO”
Low Battery Indicator
Yes “LO”
Auto Shut Off
20 minutes after last key operation
Display
Custom LCD
Operating Temperature
0 to 50ºC
Power Requirements
(4) AAA alkaline batteries (included)
Battery Life
>100 hours
Meter Dimensions / Weight
15.7 x 8.5 x 4.2 cm / 255 g