3
Calibration
12
1
If necessary, press
MODE
key to select
conductivity mode.
2
Rinse the probe with deionised water or
a rinse solution, then rinse with a small
amount of calibration standard.
NOTE: For Automatic Calibration
you must use one of the calibration
standards listed in Table 1.
3
Dip the probe into the calibration standard.
Stir the probe gently to create a homogeneous
sample. Allow time for the reading to stabilise.
4
Press
CAL
key to enter conductivity calibration
mode. The
[
CA
]
indicator will appear briefly,
then a value will appear flashing.
NOTE: To exit calibration without
confirmation, press CAL to return
to measurement mode.
5
When the value is stable, press
ENTER
. The
calibration standard value will appear for
3 seconds. If the calibration is successfully
performed,
[
donE
]
will be displayed briefly
before meter returns to measurement mode.
6
Repeat steps 1-5 as needed with additional
calibration standards.
3.4 Using Automatic Calibration (Micro 600 COND)
In Automatic Calibration mode, the COND can accept up to 4 calibration points
with maximum of 1 point per measurement range.
Note: values in the 0.00
to 20.00 µS/cm range cannot be calibrated in Auto Calibration mode.
Manual calibration is only useful when you wish to use one or more standard
values that are not listed in table 1.
See
Section 5.3 Advanced Setup
to modify automatic or manual calibration.
Measurement Mode