3
Calibration
10
The Micro 600 COND and Micro 600 TDS have five measuring ranges listed below.
Each range can be calibrated to one point per range (five total points if each
range is calibrated). Calibration is recommended for each range that will be used.
New calibrations replace old calibrations on a per range basis. For example,
if the meter has been calibrated with 1413 µS/cm (0 to 2000 µS/cm range)
and a calibration is performed with 1500 µS/cm (also 0 to 2000 µS/cm
range), the meter will replace the 1413 µS/cm calibration in that range.
The meter will retain all calibration data in other ranges.
When the electrode is replaced, it is best to clear all calibration data.
To erase all calibration data completely, see
Section 5.8 Restore Factory
Default Values
.
Micro 600 COND
3.1 Important Information on Meter Calibration
Conductivity Range
0.00
20.00 µS/cm
6.00 to 17.00 µS/cm
0.0
200.0 µS/cm
60.0 to 170.0 µS/cm
0
2000 µS/cm
600 to 1700 µS/cm
0.00
20.00 mS/cm
6.00 to 17.00 mS/cm
0.0
200.0 mS/cm
60.0 to 170.0 mS/cm
Recommended Calibration Solution Range
Micro 600 TDS
TDS Range
0.00
10.00 ppm
3.00 to 8.50 ppm
10.0
100.0 ppm
30.0 to 85.0 ppm
100
1000 ppm
300 to 850 ppm
1.00
10.00 ppt
3.00 to 8.50 ppt
10.0
200 ppt
30.0 to 170 ppt
Recommended Calibration Solution Range