5
Advanced Setup Functions
20
1
Enter advanced setup as described in Section 5.1.
2
Press
s
or
t
until
[
tdS
]
appears. Press
ENTER
.
3
Press
s
or
t
to select the desired TDS factor.
4
Press
ENTER
to select and return to the
[
tdS
]
setup function.
5
Press
s
or
t
to move to the next setup function
or press
CAL
to exit to measurement mode.
5.4 Setting the TDS Factor (Micro 600 TDS only)
The concentration of salts dissolved in solution increases the conductivity. This
relationship varies from salt to salt and is roughly linear over a given range for
a given salt. The TDS conversion factor is the number
used by the meter to convert from conductivity to TDS.
The TDS conversion factor can be set from 0.4 to 1.0.
Default value is 0.5.
See
Section 13 - Calculating TDS Conversion Factor
.
5.5 Temperature Coefficient
The temperature coefficient is the amount of change in conductivity per degree
temperature (% per °C). For best results, determine and enter the exact
temperature coefficient of your solution. The meter allows adjustment from
0.0 to 3.0% per °C. Default value is 2.1% per °C.
Note: Select 0.0% for uncompensated
measurements. Temperature can be
measured by probe and displayed.
See
Section 14 - Calculating
Temperature Coefficients
.
1
Enter advanced setup as described in Section 5.1.
2
Press
s
or
t
until
[
t.Co
%]
appears. Press
ENTER
.
3
Press
s
or
t
to select a value between 0.0 to 3.0%.
4
Press
ENTER
key to select and return to the
[
t.Co
%]
setup function.
5
Press
s
or
t
to move to the next setup
Pr
es
s
EN
TE
R
t
o
co
nf
irm
s
el
ec
tio
n;
o
r
pr
es
s
C
A
L
to
e
sc
ap
e
w
ith
ou
t
co
nf
irm
in
g
se
le
ct
io
n
Pr
es
s
EN
TE
R
t
o
co
nf
irm
s
el
ec
tio
n;
o
r
pr
es
s
C
A
L
to
e
sc
ap
e
w
ith
ou
t
co
nf
irm
in
g
se
le
ct
io
n