O
PERATION
M
ANUAL
VKL
10
/100
V
ERSION
V005-06-2020
26
The particle-technical function test of the VKL series can be accomplished so simply on-site
by everyone. At the same time, this test proves that the measurement 1 in position 1 was
coincidence-free.
Please note:
If the dilution factor determined in such a way is not correct, then either the
measurement was still performed in coincidence in position 1 during measurement 1, or the
test system must be maintained.
Calculation of the laboratory air concentration:
Formula 3: Calculation of the laboratory air concentration
100
min
28300
min
228016
3
=
cm
particle
C
labair
3
806
cm
particle
C
labair
=
Please note:
This formula may be only used, if by the measurement 2 it was clearly shown
that the measurement 1 was coincidence-free performed. Without the proof of the
coincidence-free measurement in position 1, the determination of the laboratory air
concentration is extremely doubtful!