5. Equipment Layout
5.1. The Pneumatic System
Figure 5.1. shows the pneumatic system. Pre-cleaned compressed air is conducted via a
toggle switch to a pressure reduction valve. Via a permanently adjusted throttle and the pre-
pressure indicated at the pressure gauge, the volume flow rate may be adjusted
reproducibly. A filter guarantees sufficiently particle-free air. This particle-free air flows
through the annular nozzle and causes the sucking-in of the aerosol to be diluted. Via the
outlet nozzle, a portion of the diluted aerosol is isokinetically sucked out for further
treatment. The excess aerosol is conducted outside through the excess air outlet.
throttle
filter
pressurized air
toggle switch
reduction valve
pressure gauge
excess air outlet
mixing chamber
inlet nozzle
excess air tube connector
outlet nozzle
Fig. 5.1: Pneumatic control panel