Stimulation is indicated by synchronous flashing of ‘PACE’ & Sensing is indicated by
synchronous flashing of ‘SENS’. Mains independence battery operation provides a maximum
on electric safety for both patient and user.
4.1
Technical Data
PARAMETERS
VALUES
Technical Parameters
Description of
Pacemaker
Dual chamber External Pacemaker
Rate
30 to 250 BPM
(
In steps of 2 BPM)
Upper Rate
Base Rate + 30 BPM (Min 110 BPM)
Rapid Atrial Pacing
60 to 1000 BPM
Amplitude (A)
0.1 TO 18 Volts (at 510
Ω
load)
Amplitude (V)
0.1 TO 18 Volts (at 510
Ω
load)
Sensitivity
ATRIUM 0.4 To 10 mV
VENTRICLE 0.8 To 20 mV
Mode
AOO,AAI, DOO, DVI,DDI, DDD,VDD ,VVI, VOO
AV Delay
20 to 300 ms
Pulse Width
1.0 -- 1.5 ms
V Blanking Period
125 ms(Pace),75ms(Sense)
Refractory (PVARP in
DDD)
Atrium 150-500ms, Ventricle 250ms
Battery Life
166 hrs
Battery Chemistry
Duracell 9V PP3 Alkaline battery
Polarity
Bipolar
Indication
Pacing Indication :
Flashing RED LED’s on the front panel for both Atrial
& Ventricle.
Sense Indication :