![Pacetronix ZEUS-1 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/pacetronix/zeus-1/zeus-1_manual_3116798015.webp)
Description of Blanking period :
Blanking is the period
following a paced Atrial event for which Ventricle sense
amplifier will be kept ‘OFF’.
5.2.7
‘V’ Output (Volts) (Ventricle Pulse Amplitude)
This knob is used to set Ventricle pulse Amplitude in volts
values. It allows continuous adjustment of the amplitude
from 0.1 to 18 Volts.
Description of Ventricle Pulse Amplitude:--
It is the
electrical pressure that causes electric current to flow.
5.2.8
‘V’ Sens (mV) (Ventricle Sensitivity)
This knob is to select different Ventricle sensitivity values.
It allows continuous adjustment of the Sensitivity from 0.8
to 20 mV.
Description of Ventricle Sensitivity :--
Sensitivity is the
ability of the pacemaker to see intrinsic cardiac events (R
Wave).
5.2.9
Blanking Period
Blanking is the period following a paced Atrial event for which Ventricle sense amplifier will
be kept ‘OFF’. Blanking period is fixed For Pace- 125ms and for Sense-75ms.
5.2.10
Rapid Atrial Rate
Caution:
RAP is for atrial use only. Be sure the leads are connected to the atrium, not the
ventricle, before enabling RAP.
Rapid atrial rate is used to control Atrial fibrillation. Rapid atrial rate can be activated by
pressing RAP button located on front panel of the unit and by selecting from rate setting
knob.
Caution:
RAP may result in tachycardia, acceleration of existing tachycardia, or fibrillation.
Apply high rates under careful patient monitoring and control. Monitor the patient’s ECG and
Содержание ZEUS-1
Страница 7: ...ZEUS 1 ...