Defibrillation / Cardioversion
Defibrillation discharges of 360 watt-seconds have not affected the Model Zeus-1
in laboratory tests. However, for maximum safety it is recommended that the
paddles not be placed near the Model Zeus-1 or the lead system. Whenever
possible, for the safety of the patient, disconnect the pacemaker from the lead
system before defibrillating or cardioverting. A relatively low resistance pathway
exists between the positive (+) and negative (-) electrodes of the implanted lead
system. During defibrillation a large current could flow across this pathway,
causing myocardial damage.
Connecting the Lead System
The patient cable should be connected to the temporary pacemaker before the lead
system is connected to the patient cable.
Turning the Device On
Patient lead, cable and device connections should be made before the pacemaker
is powered on .Before connecting to patient parameters should be verified.