![Pacetronix ZEUS-1 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/pacetronix/zeus-1/zeus-1_manual_3116798019.webp)
and External pacemaker (Zeus-1) as described in the method below.
Method : Twist the terminal caps counter clockwise, push the lead
electrodes into the end of the collets terminals, tighten the leads by
twisting caps clockwise.
Caution : The terminal caps should not be over tightened otherwise the
terminal may break.
6.10
Battery Replacement
It is recommended to replace battery when Low battery Indicator
LED glows.
In the Low Battery case the unit will continue to pace normally, on the
last set parameters but the battery should be replaced as soon as
possible.
Under normal conditions at nominal values of continuous operation, a
9V alkaline battery will have a service life of approximately 166 hours
(7 days) on full battery.
Nominal Values: Mode- DDD , Rate – 70 BPM , Amp (A/V) – 5.0V
AVI – 140 ms, Sens (A/V) – 0.8mV/2.0mV
NOTE: We recommend having a spare battery all the times. The
battery should be removed if the device is placed in storage.
Battery replacement time is approx 15 seconds.
Содержание ZEUS-1