37
DIAGNOSTICS
Note:
SHARK is not user serviceable. Specialized tools are necessary for the repair of any SHARK component.
INTRODUCTION
A flashing SHARK information gauge indicates there is an abnormal condition somewhere on the powerchair.
The components that SHARK provides fault information for include, the motor, the park brakes, the batteries,
the cabling and the SHARK modules themselves.
Note that joystick OONAPU (Out OF Neutral At Power Up) is not a fault. Simply by removing your hand from
the joystick and allowing it to return to the neutral position, the fault will immediately clear.
The nature of the abnormal condition is indicated by a flash code. This is a sequence of flashes, separated by
a pause, followed by a repetition of the sequence. The number of flashes relates to the condition. For instance,
four flashes of the SHARK information gauge, a pause, followed by four flashes, etc. indicates a right motor
fault. Five flashes would indicate a left park brake fault.
Depending on the severity of the condition, the powerchair may or may not allow driving. In some cases the
chair may be allowed to drive but in a reduced speed ("limp") mode.
DIAGNOSTICS & FAULTS
Flash codes indicate the nature of an abnormal condition directly from the
SHARK Information Gauge. Without the use of any servicing tools, the condition
can be simply diagnosed.
1
User Fault
Possible stall timeout or user error.
Release the joystick to neutral and try again.
2
Battery Fault
Check the batteries and cabling.
Try
charging
the
batteries.
Batteries
may
require
replacing.
3
Left Motor Fault
Check the left motor, connections and cabling.
4
Right Motor Fault
Check the right motor, connections and cabling.
5
Left Park Brake Fault
Check the left park brake, connections and cabling.
6
Right Park Brake Fault
Check the right park brake, connections and cabling.
7
SHARK
Check the SHARK Communications Bus connections and wiring
Joystick Console Fault
Replace the Joystick Console.
8
SHARK
Check SHARK connections and wiring
Power Module Fault
Replace the Power Module.
9
SHARK
Check Battery voltage is greater than 17V.
Communications Fault
Check the SHARK Bus cable.
Replace
the
SHARK
Power
Module.
Replace
the
SHARK
Joystick
Console.
10
Unknown Fault
Check all connections and wiring.
11
Incompatible
Wrong type of Control Unit connected.
System Components
Ensure the branding of the Power Module matches the Joystick Console.
CODE
DESCRIPTION
CORRECTION
If the light continues to flash after following the procedures in the LED Diagnostic Code chart, do
not ride the wheelchair. Notify your dealer for repair.
Содержание Scout M Series
Страница 2: ...48 SPECIAL OPTION WARNINGS ...