3.3.2 Installation - Micrometer
1. For easier installation, retract
micrometer adjustment arm - do
not force movement beyond limits
as this may damage the
micrometer.
2. Loosen bolts on sample support
arm and slide smaller diameter side
of micrometer shaft into outer shaft
guide. DO NOT tighten bolts on
sample support arm at this point.
3. Push micrometer into the outer shaft
guide thus pushing the micrometer
shaft to the right. While holding
micrometer in place, tighten set
screw on outer shaft guide. NOTE:
The hole in the micrometer does not
need to line up with the bolt.
DO NOT over tighten the bolt as this
may damage and pinch the
micrometer.
4. Attach sample fixture to sample
support arm and then align holder
to approximately where the
wafering blade will be installed.
Tighten sample holding arm bolts.
Retract micrometer arm
Slide smaller end of
micrometer shaft into outer
shaft guide
Tighten set screw:
DO NOT OVER
TIGHTEN as this
may damage
micrometer
Install sample
fixture, align
sample support
arm and then
tighten bolts