background image

25 

 

7

 

Product specification 

7.1

 

Main unit 

Electrical Safety classification 

Class II, BF type applied part; non 
AP/APG equipment 

Liquid protection level 

IPX0 

Operating mode 

Continuous 

Product life 

5 years 

Atomizing port pressure range of 
concentrator with atomizing 

60kPa - 250kPa 

Atomization outlet pressure 

30kPa -70kPa 

Fuse 

F5AL250V 

Power supply 

AC 220V±22V, 50±1Hz 

 

7.2

 

Environmental Specifications 

Operating 
environment 

Temperature: 5°C - 40°C 

Relative humidity: ≤75% 

Barometric pressure: 86kPa - 106kPa 

Storage and 
transportation 
environment 

Temperature: -40°C - +55°C 

Relative humidity: ≤93% 

Barometric pressure: 50kPa - 106kPa 

 

7.3

 

Concentrator models and Technical parameters 

Below table shows concentrator models and main technical parameters. 

 

Type 

Rated 

flow 

(L/min) 

Concentration 

(V/V) 

Noise 
(dBA) 

Size (mm) 

(LxWxH) 

±20mm 

Power 

consumption 

(VA) 

Weight 

(kg) 

±3kg 

Atomization 

(mL/min) 

KSN-3   

 

 
 
 

 

93%±3% 

 

 

≤53 

 
 
 

 

410X280X510 

 

 

≤320 

 

 

18 

None 

KSN-
3N 

≥0.2 

KSN-3T 

None 

KSN-
3M 

≥0.2 

KSN-5   

 

 

 

≤54 

 

 

≤480 

 

 

20 

None 

KSN-
5N 

≥0.2 

KSN-5T 

None 

Содержание Pure LN

Страница 1: ...O2 Read this manual carefully before use Ozocheck Pure LN Oxygen Concentrator User s Manual O2 O2 O2 2 O2 O2 O2 O2 2 O2 O2 O2 O2 O2 O2 O2 O2 2 O O O...

Страница 2: ...e photocopied reproduced or translated into other languages Materials protected by the copyright law including but not limited to confidential information such as technical information and patent info...

Страница 3: ...Medsource may provide with compensation necessary circuit diagrams and other information to help qualified technician to maintain and repair some parts which Medsource may define as user serviceable...

Страница 4: ...ion 9 3 5 Power ON OFF 12 3 6 Oxygen uptake operation 12 3 7 Atomization operation of N M type oxygen concentrator not administered 14 3 8 Other operations 14 3 8 1 Timing operation 14 3 8 2 Remote co...

Страница 5: ...s and Technical parameters 25 8 Warranty and Service 28 8 1 Warranty 28 8 2 Contact information 29 9 EMC 30 9 1 Electromagnetic emissions 30 9 2 Electromagnetic immunity 31 9 3 Electromagnetic immunit...

Страница 6: ...oor physique pregnant women college entrance examination students and it also be used to eliminate fatigue and restore the body function after heavy physical or mental consumption Concentrator s safet...

Страница 7: ...t shows the electrical control process 1 3 Equipment symbols Some symbols may not appear on your equipment Symbol Description Symbol Description Note Consult the accompanying documents Do not expose t...

Страница 8: ...his side up Fuse ON power supply Class II equipment OFF power supply Alternating current Type BF applied part Manufacture Date Consult instructions for use Serial number Increase or decrease Knob Batc...

Страница 9: ...the plug cord from the power outlet if long time no use of concentrator Do not pull out the power cord when removing the plug 2 1 Safety tips for oxygenconcentrator Electrical shock hazard do not dis...

Страница 10: ...near high frequency disturbing equipment such as speaker MRI or CT etc Do not use parts accessories or adapters that are not approved by the manufacturer Using other humidifier or other accessories n...

Страница 11: ...rent power outlet Please pay attention to electricity safety Do not turn on the concentrator if the plug or power lines damaged Ensure to cut off the power when cleaning the concentrator or cleaning a...

Страница 12: ...ified by our company The atomization device and the nasal cannula as accessories provided by our company is the test sample The size of the atomization cup is 6 12mL and the length of the nasal cannul...

Страница 13: ...tor The unit should be stored in the environment without corrosive air and with a good ventilation Be careful to transport or convey don t let it reversal or thwart tilt angle not greater than 5 NOTE...

Страница 14: ...xygen tube This port is connected to oxygen outlet Humidifier bottle Handle Caster Air outlet 4 5 Front panel 1 2 3 1 Indicator LED Power indicator This indicator lights green when the oxygen concentr...

Страница 15: ...s with buzzer alarm tone This can prompt the user to disconnect the power and check the power supply system 2 Start Stop oxygen button 3 Timer timing increase decrease button Adjust the timing increas...

Страница 16: ...PTFE Provides the air pressure necessary for adsorption and used to separate oxygen from air Air pretreatment Foam filter ABS resin nonwovens Used to provide gas cooling water removal filtration etc C...

Страница 17: ...ing operation After use touch the ON STANDBY button to stop oxygen supply and disconnect the tube or connecting pipeline from the oxygen outlet then press the I O power switch to position O OFF Unplug...

Страница 18: ...ion for easy inhale of the pure oxygen and get maximum comfort The oxygen density could be reached to 90 within 12 minutes 9 Turn off the equipment after use Pull out the power pin and take care of th...

Страница 19: ...tomizing Remove tubing and turn off the oxygen concentrator by pressing the I O power switch Atomization performance The atomization outlet pressure is 60kPa to 250kPa and the maximum air flow rate is...

Страница 20: ...he remote controller cannot work the battery may not have sufficient power Please replace the battery periodically with the same battery and be careful of the affirmative and negativepole NOTE The con...

Страница 21: ...age and cleaning of Headset Headset can be worn on head it makes user comfortable with its excellent personal feeling Unique perfuming function add different essential oil into liquid cup this aromath...

Страница 22: ...uid cup top cover OPEN CLOSE Liquid cup base Cleaning Rinse or immerse the disassembling liquid cup in warm water for about 15 minutes Add appropriate amount of vinegar in the water for cleaning more...

Страница 23: ...on Beep continuous sounding Flashing frequency of the red alarm lamp 1 4Hz 2 8Hz High Please turn off the power immediately If an alarm still exists after confirming that the power supply and connecti...

Страница 24: ...t the plug to thesocket correctly 2 Check the power supply 3 Replace the fuse After turned on the switch the light works but the oxygen concentrator doesn t work 1 Air compressor protection 2 Inlet or...

Страница 25: ...Description E01 The pressure cannot reach the required for 25s or longer after startup E02 During operation the pressure suddenly drops below the limit pressure 20kPa E05 The pressure exceeds the lim...

Страница 26: ...e equipment from the power supply before performing maintenance to avoid electric shock Person without training or authorization of manufacturer cannot open thecabinet 5 1 Care and Cleaning of cabinet...

Страница 27: ...stallation or with wet sponge Otherwise it will damage the oxygen concentrator Disassembly The filter sponge is located on the back of the oxygen concentrator as shown below pull the filter out and re...

Страница 28: ...power failure alarm occurs you can use the tool to remove the screw and replace the fuse before restarting the concentrator 5 5 Disposal of waste Dispose of the waste disposable nasal cannula filter...

Страница 29: ...on cannot be guaranteed and the patient may be injured Check the accessories and their packages for any sign of damage Do not use them if any damage is detected The accessory material that contacts th...

Страница 30: ...ifications Operating environment Temperature 5 C 40 C Relative humidity 75 Barometric pressure 86kPa 106kPa Storage and transportation environment Temperature 40 C 55 C Relative humidity 93 Barometric...

Страница 31: ...26 KSN 5M 0 2...

Страница 32: ...oncentration delivered by oxygen concentrator Altitude m 800 800 1500 1500 3000 3000 4000 KSN 3 series 90 82 75 65 KSN 5 series 90 80 72 63 Impact of outlet pressure on output flow When the outlet nom...

Страница 33: ...28 Rated flow 100...

Страница 34: ...29 Applicable condition for the indicator of oxygen concentrator Temperature 5 40 C Relative humidity 80 Barometric pressure 86kPa 106kPa...

Страница 35: ...ipment cannot be used within 12 month after sold More than 12 month the company provides the parts to repair with reasonable charge if it could not be used The following conditions are not covered by...

Страница 36: ...for reference only Please take the real product you receive as standard No notice will be made in case of any update 8 2 Contact information If any questions in equipment operation please contact the...

Страница 37: ...y result in increased emissions of decreased immunity of the Oxygen Concentrator The Oxygen Concentrator should not be used adjacent to or stacked with other equipment If adjacent or stacked use is un...

Страница 38: ...30 Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 1 kV for input output lines 2kV for power supply lines Not applied Mains power quality should be that of a typical commerci...

Страница 39: ...electromagnetic environment specified below The customer or the user of the concentrator should assure that it is used in such an environment As following Immunity test IEC60601 Test level Compliance...

Страница 40: ...dio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey sho...

Страница 41: ...38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equatio...