27
Remarks:
Output oxygen concentration range for KSN-3 series
concentrator: 3L/min; 93%±3%.
Output oxygen concentration range for KSN-5 series
concentrator: 5L/min; 93%±3%.
Model with “N” is feature configuration, which is added atomization
function on the basic functions.
Model with “T” is feature configuration, which is added the function of
remote control.
Model with “M” is feature configuration, which is added the functions
of atomization and remote control.
The effect of altitude on oxygen concentration delivered by oxygen
concentrator:
Altitude(m)
≤800
800 - 1500
1500 - 3000
3000- 4000
KSN-3
series
≥90%
≥82%
≥75%
≥65%
KSN-5
series
≥90%
≥80%
≥72%
≥63%
Impact of outlet pressure on output flow:
When the outlet nominal pressure is 0kPa, the specified control flow value
for KSN-3 series concentrator is 3L/min; Output oxygen concentration range
for KSN-5 series concentrator is 5L/min.
When the outlet nominal pressure is 7kPa, the specified control flow value
for KSN-3 series concentrator is 2.8L/min; Output oxygen concentration
range for KSN-5 series concentrator is 4.5L/min.
Function between Oxygen concentration of concentrator and rated flow
(outlet nominal pressure is 0kPa)
Concentration
(V/V%)