11
Display screen
Timer timing
Concentration
Flow value Total running time
Remote controller
1.
Working indicator (the indicator lights green when using buttons of
remote controller)
2.
Timing 1 hour button
3.
Start/Stop oxygen button
4.
Timing 2 hours button
5.
Timing 3 hours button
Main structure and material of concentrator
Main Structure
Material
Description
Air compressor
ZL102 cast
alum, filling
PTFE
Provides the air pressure necessary for
adsorption and used to separate oxygen
from air
Air pretreatment Foam filter,
ABS resin,
nonwovens
Used to provide gas cooling, water
removal, filtration, etc.
Control valve
/
Control the compressed air treated by
the air pretreatment system into the
molecular sieve adsorption tower for
periodic pressurization and exhaust
Molecular sieve
adsorption tower
6063
aluminum
alloy,
molecular
sieve
The molecular sieve is filled in a closed
container. Oxygen in the air is separated
by the characteristic of selective
adsorption of gas by a molecular sieve.
1
2
3
4
5