13
3.
Install the humidification bottle on the humidifier bottle holder of
concentrator and secure it with band.
4.
Press the power switch to position “I”(ON). The power is plug in when
the power indicator turns green. The oxygen concentrator is in working
state.
5.
Connect the humidification bottle to the oxygen outlet using a
connecting tube and connect the other port of the humidification
bottle to the nasal oxygen tube (as shown below).
This port is
connected to
the oxygen
outlet
This port is
connected to
nasal oxygen
tube
Maximum level
Water level
Minimum level
6.
Wear a nasal oxygen tube.
7.
Press the button
on the front panel to start oxygen supply.
8.
Adjust the cannula to appropriate position for easy inhale of the pure
oxygen and get maximum comfort. The oxygen density could be
reached to 90% within 12 minutes.
9.
Turn off the equipment after use. Pull out the power pin and take care
of the cannula for another use.
Every few seconds the concentrator issues "TAPOO" sound while operating,
it is the normal ventilation sound.
Choose safety-qualified socket and socket board with safety electrical
certification.
Do not remove the cabinet unless qualified technician, in case of
damaging the machine, electric shock or other accident.
Oxygen uptake time and oxygen flow adjustment should follow the
physician's advice.
Check that the tubing or accessory is clogged, kinked or the humidifier is
damaged if the flow value is lower than 0.5L/min.
If no use of concentrator for a long-term, the water in humidifier bottle
should be drained. Save the humidifier bottle after cleaned and dried.
WARNING