5.Advanced User Guidebook
(4)
Set the network parameters of the oscilloscope
. In the oscilloscope, press the
Utility
button and press
H1
button, turn the
M
knob to select the
LAN Set
; press
the
H2
button, the set menu is displayed on the right. Set the
IP
and the
Port
to
the same value as the "Ports-settings" in the software in step (3). Press the
H3
button to select "
Save set
", it prompts "reset to update the config". After resetting
the oscilloscope, if you can get data normally in the oscilloscope software, the
connection is successful.
Figure 5-52 Set the network parameters of the oscilloscope
How to connect with the computer through a router:
(1)
Connection
. Use a LAN line to connect the oscilloscope with a router, the LAN
port of the oscilloscope is in the right side panel; the computer should be
connected to the router too.
(2)
Set the network parameters of the computer
. Since the oscilloscope can not
support obtaining an IP address automatically, you should assign a static IP
address. The Default gateway and Subnet mask should be set according to the
router. Here we set the IP address to 192.168.1.71, Subnet mask is 255.255.255.0,
Default gateway is 192.168.1.1.
62