5.Advanced User Guidebook
OFF
,
1second
,
2second, 5second
and
Infinity
.
When the "
Infinity
" option is set for
Persist
time
,
the measuring points will be stored till the controlling value is changed
(see
Figure 5-34
). By pressing the
F2
button, the persistence will be cleared.
Figure 5-34 Infinite Persistence Display
XY Format
This format is only applicable to Channel 1 and Channel 2. After the XY display
format is selected, Channel 1 is displayed in the horizontal axis and Channel 2 in the
vertical axis; the oscilloscope is set in the un-triggered sample mode: the data are
displayed as bright spots.
The operations of all control knobs are as follows:
The Vertical
VOLTS/DIV
and the
VERTICAL POSITION
knobs of Channel 1
are used to set the horizontal scale and position.
The Vertical
VOLTS/DIV
and the
VERTICAL POSITION
knobs
of Channel 2
are used to set the vertical scale and position continuously.
The following functions can not work in the XY Format:
Reference or digital wave form
Cursor
Time base control
Trigger control
FFT
47