![Oval ULTRA OVAL Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/oval/ultra-oval/ultra-oval_instructions-manual_1662168026.webp)
B-144-12-E
26
11.1.1 Function Setup Switch "SW2"
NOTE:
shows default setting.
CAUTION:
To change SW1 settings, place all switches in OFF and then place only the requied
switch in ON.
11.1.2 Output Mode Setup Switch "SW1"
DIP Switch No.
SW2
−
1
SW2
−
2
SW2
−
3
SW2
−
4
SW2
−
5
SW2
−
6
Description of Functions
Parameter write-protect
OFF: Rewritable "Default"
ON: Write protect
* To seal, set to ON before pasting the seal.
Factored/unfactored pulse select
OFF: Factored pulse
ON: Unfactored pulse
Inhibit to reset the resettable total
OFF: Resettable "Default"
ON: Nonresettable
Accumulated total reset
ON
→
OFF resets the accumulated total.
* Set to OFF in normal use
.
Unused
* Always OFF
Battery power switch
OFF: Battery power OFF
ON: Battery power ON "Default"
Output Mode
DIP
Switch No.
SW1
−
1
SW1
−
2
SW1
−
3
SW1
−
4
SW1
−
5
SW1
−
6
SW1
−
7
SW1
−
8
SW1
−
9
SW1
−
10
Remarks
ON for voltage pulse only,
All else OFF.
ON for voltage pulse/analog output only.
All else OFF.
ON for open collector output and voltage
pulse output only.
All else OFF
Power pulse/
analog
OFF
OFF
ON
ON
ON
ON
OFF
OFF
OFF
OFF
Open collector
pulse
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
Voltage pulse
ON
ON
OFF
OFF
OFF
OFF
ON
ON
ON
ON