10. DISASSEMBLY AND INSPECTION
◎
Although it depends on individual operating conditions, periodic disassembly and
inspection should be performed at least once a year.
CAUTION: Because Ultra Oval is a precision industrial instrument, disassembly
and inspection should be performed indoors as a rule. If it is desired to
disassemble and inspect it as installed in the field, an important precaution
to remember is to reduce the internal pressure of the piping assembly to a
safe level, shut off valves upstream and downstream of the meter fully, drain
the piping assembly and then place a suitable fluid receptacle directly below
the Ultra Oval. Exercise care to keep individual members disassembled free
from grit and dust.
CAUTION: If the meter is of externally powered type, be sure to turn off power to the
register before you disassemble.
◆
Meter Body Inspection
◆
Complaint and possible cause: Process fluid will not run.
Jammed with scales, the Oval rotors fail to rotate, resulting
in an interrupted fluid flow.
10.1 Size 50 Meter Body Disassembly and Inspection
①
Take off all six hex bolts (104) (M6), using a
wrench (10mm).
With hex bolts removed, the register assembly
is ready to be separated from the meter body.
Be careful not to drop the register assembly.
②
With the front cover (register end) removed,
the measuring chamber with Oval rotors and
other components are now accessible.
Hex Socket
Head Bolt
Meter Body
Front Cover
(register end)
Hex Socket
Head Bolts
O-Ring