14. MAINTENANCE
As long as your DELTA meter is placed in service under proper operating conditions, it will perform
satisfactorily over extended periods of time. But scale, sludge and other foreign matter contained in the
metered fluid could adhere to the inner walls of your meter and degrade meter performance in time. For
this reason, periodic disassembly and inspection is suggested.
14.1 Disassembly and Inspection
◇
Time for disassembly and inspection
◇
() Six months after the first time operation.
(2) According to the conditions observed in () above, determine the second time disassembly and
inspection. (Once a year is suggested under normally encountered operating conditions.)
◇
Inspection points
◇
() Probe (probe body and bluff body) for deposits.
(2) Observe the output waveform from the amplifier board in the amplifier and make sure that it conforms
to the waveform shown in the section under the topic "TROUBLESHOOTING" in the preamplifier
instruction manual.
14.2 Disassembly and Inspection
NOTE:
See Section 5, "EXPLODED VIEW AND PARTS LIST, FIXED TYPE" on pages and 2.
(1) Probe Inspection
Taking off hex bolt (0) and three hex socket head setscrews (0A), the probe body (0) can be
separated. Inspect probe body (0) on its inner walls and bluff body (05) for condition and clean.
Foreign matter in the capillary slots (a total of four places) on the sides of bluff body in particular must be
removed thoroughly. Adhere to the following instructions during this procedure:
CAUTION
①
Do not force the bluff body because it contains sensor (201) installed and accessible
through capillary slots.
②
Sharp edges of the bluff body play an important role for maintaining meter accuracy.
Exercise care not to damage them.
Following inspection, install in place. Align the locating pin on the probe body (101) with
the mating pin slot in bluff body (105) at this time.
(2) Sensor Removal
If it is desired to remove the sensor for sensor replacement or other reason, proceed as follows:
①
Firstly separate the preamplifier in the following manner:
(a) Remove power.
(b) Remove cover (30) on the terminal box. (
→
see page 23).
NOTE: With explosionproof spec., separate lcok (30) before removing the cover.
(
→
see page 23).
(c) Disconnect the cable of external wiring. (
→
see page ).
(d) Separate shield strip (30). (
→
see page ).
(e) Disconnect sensor leads. (
→
see page ).
(f) Loosen hex socket head setscrews (30) and pull out the preamplifier upward.